EL3000 series is a dot printing type analog recorder sized 144x144mm with 100mm width chart. The unit starts recording as soon as the power supply and input are connected and it is also easy to operate. Scale plate, input range and function of the recorder can be selected for various purpose and applications as many kinds of options are prepared.
Universal power supply Universal power supply with voltage range of 100 to 240V AC (50/60Hz) is applied. |
Linear temperature scale The accuracy is ±0.05% and the resolution is 1ƒÊV or 0.1°C |
Various industrial values can be measured at the same time with selectable ranges Temperature scale of thermocouple and resistance thermometer input is a linear scale that is excellent in reading indication value. |
Standard 6 chart speeds 6 chart speeds (5,10,20,40,80,160mm/h) are switchable as standard. 5 chart speed and hour/minute change are prepared as option. |
Alarm setting (common alarm) as standard Higher and lower limit alarm can be programmed for every point. Alarm value is easy to be programmed by pointer location. You can check the alarm by front LED lighting. Alarm output is prepared as option. |
CE marking The EL recorder is conformed to the rules of safety standards of CE. (Conformity pending) |
Unit structure and light-weight Light-weight (50% of the previous unit weight) was realized by easy maintenance unit structure. |
Employing removable type terminal board Employing easy connecting removable type terminal board. |
Multi-Point Type
Thermocouple | ||
Model | Specification | Remarks |
EL3D15-NNN | 1 point | Accessories |
EL3D25-NNN | 2 points | Accessories |
EL3D35-NNN | 3 points | Accessories |
EL3D65-NNN | 6 points | Accessories |
RTD *1 | ||
EL3D17-NNN | 1 point | Accessories |
EL3D27-NNN | 2 points | Accessories |
EL3D37-NNN | 3 points | Accessories |
EL3D67-NNN | 6 points | Accessories |
*1 | Optional built-in voltage divider and thermocouple burnout input is only type "7" |
Option |
Model | Specification |
EL3D** - ⬚:NN | w/Input and scale plate |
⬚: |
0: Standard input + standard scale plate 1: Non-standard input (including current input and built-in voltage divider) + standard scale 2: Standard input + non-standard scale plate 3: Non-standard input (including current input and built-in voltage divider) + Non-standard scale |
*2 |
Double scale is available. Input and scale selection are needed for non-standard input and non-standard scale plate. |
Model | Specification |
EL3D** - N⬚N | Alarm |
⬚: |
0: None 1: 2 alarm outputs |
Model | Specification |
EL3D*7 - NN⬚ | Chart speed + burnout |
⬚: |
0: Standard 6-speed + burnout disabled 1: Standard 6-speed + up-scale burnout 2: Standard 6-speed + down-scale burnout A: Standard 5-speed hour/minute change + burnout disabled B: Standard 5-speed hour/minute change + up-scale burnout C: Standard 5-speed hour/minute change + down-scale burnout |
*3 | Burnout on all channels is programmed together for thermocouple/residence thermometer input. |
Input Specifications | |
Measuring points: | 1,2,3 and 6 points |
Reference range and types: |
DC voltage: ±13.8mV, ±27.6mV, ±69mV, 200mV, ±500mV,±2V, ±5V Built-in voltage divider; ±10V, ±25V, ±50VD Ccurrent: External installation of shunt resistor(250Ω) is applied (option) Thermocouples: K, E, J, T, R, and B (option) Resistance thermometer: Pt100(1997) (Measured current; 1mA)* Linear scale for thermocouple and resistance thermometer |
Input designation: | Single scale (standard), double scale (option) |
Accuracy rating: | ±0.5% of input span (except for some input under standard operating condition) Refer to the table of standard range and minimum width of scale for non-standard input |
Indicating deadband: | 0.3% of input span |
Reference junction compensation accuracy: |
K,E,J,T --- ±1.0°C or less (23°C±10°C) ±2.0°C or less (0 to 50°C) (For internal reference junction compensation, the errors above are added to the accuracy rating) |
Temperature drift: | ±0.02%/°C (Converted into reference ranges) |
Measurement cycle: | 6 seconds/point |
Indicating resolution: | Approximately 1/2,000 |
Burnout (option): |
On thermocouple or resistance thermometer input, disconnection of signal can be detected .(Up-scale and down scale burnout on all channels can be programmed.) Burnout detection: Voltage application method (approximately 8V, 1mA) |
Allowable signal source resistance: |
Thermocouple inputs, DC voltage inputs (±5V or less) --- 1kΩ(burnout disabled) or less DC Voltage inputs (input more than ±5V) --- 100Ω or less Resistance thermometer inputs --- per wire 10Ω or less (Same resistance for 3 wires) |
Maximum input voltage: |
Thermocouple inputs, DC voltage inputs --- ±10V DC or less DC voltage inputs (Voltage divider built-in) --- ±60V DC or less Resistance thermometer --- ±6V DC or less |
Maximum common mode voltage: | 30V AC |
Common mode rejection ratio: | 120dB or more (50/60Hz±0.1%) |
Normal mode rejection ratio: | 50dB or more (50/60Hz±0.1%) |
Recording Specifications | |
Recording accuracy: | ±0.5% of recording span |
Recording system: | Inkpad dotting |
Balancing time: | Input span movement --- approximately 2 seconds |
Recording color: | 1: red 2: blue 3: green 4: violet 5: purple 6:brown |
Chart paper: | Fan-fold type: total width of 114mm, total length of 10m, effective chart width of 100mm |
Chart speed: | 6-speed change, 5,10,20,40,80,160mm/h (standard) |
Chart speed accuracy: | ±0.1% or less (It is based on the chart scale.) |
Indicating Specifications | |
Analog indication: | Scale plate and pointer |
Scale plate: | Single scale or double scale (minimum scale division: 80) |
Alarm Specifications | |
Alarm display: |
Pointer and alarm-point seal pasted on scale. Alarm LED lamp lightens for alarming (All channels OR output) |
Alarm types: | Higher and lower-limit alarm |
Alarm programming: |
Individual setting for higher and lower-limit value (Programming percentage of input span by indicating pointer, input resolution 0.5%) |
Alarm deadband: | 0.4% of input span |
Alarm output (option): |
1a contact and 2 outputs (common) Maximum contact capacity: 2A (resistive load), 0.5A (inductive load) |
General Specifications | |
Rated power voltage: | 100 to 240V AC, 50/60Hz (Universal power supply) with power supply switch |
Power consumption: | Maximum 12VA (100V AC), Maximum 20VA (240V AC) |
Environmental conditions: |
Reference operation condition Ambient temperature range: 21 to 25°C Ambient humidity range: 45 to 65%RH Power voltage: 100V AC±1% Power frequency: 50/60Hz±0.5% Attitude: left/right 0°, forward tilting 0°, backward tilting 0° Warm-up time: longer than 30 minutes Normal operation condition Ambient temperature range: 0 to 50°C Ambient humidity range: 20 to 80%RH Power voltage: 90 to 264V AC Power frequency: 50/60Hz ±2% Attitude: left/right 0 to 10°, forward tilting 0°, backward tilting 0 to 20° Transportation condition (at the packed condition on shipment from our factory) Ambient temperature range: -20 to 60°C Ambient humidity range: 5 to 90%RH (No dew condensation) Vibration: 10 to 60Hz, 4.9m/ S2 (0.5G) or less Impact: 392m/S2 (40G) or less Storage condition Ambient temperature range: -20 to 60°C Ambient humidity range: 5 to 90%RH (No dew condensation) |
Insulation resistance: |
Secondary terminals and protective conductor terminals --- 20MΩ or more at 500V DC Primary terminals and protective conductor terminals --- 20MΩ or more at 500V DC Primary and secondary terminals --- 20MΩ or more at 500V DC Note:Primary terminals --- Power (L,N), Alarm terminals (mechanical relay) Secondary terminals --- Measurement input terminals |
Dielectric strength: |
Secondary terminals and protective conductor terminals --- 1 minute at 500V AC Primary terminals and protective conductor terminals --- 1 minute at 1500V AC Primary and secondary terminals --- 1 minute at 2300V AC Notes; Primary terminals --- Power(L,N), Alarm terminals (mechanical relay) Secondary terminals --- Measurement input terminals |
Case: | Door (frame) --- ABS resin, window --- glass Case --- ABS resin |
Color: | Door (frame) --- Black (equivalent to Mussel N1.5),window --- Transparent Case --- Black (equivalent to Mussel N1.5) |
Mounting: | Panel mounting |
Weight: | Approximately 1.7kg |
Power voltage fluctuation: | Indication fluctuation 0.2% or less (conversed into reference ranges at 90 to 264V AC) |
Optional Specifications | |
Alarm output |
Alarm contact output is available Alarm relay --- Mechanical relay 1a contact, 2 outputs (common) Maximum contact rating: 250V AC 2A, 30V DC 2A (resistive load), 250V AC 0.5A, 30V DC 0.5A (inductive load) |
DC current input | 250Ω of shunt resistor is applied to measure voltage input |
Built-in voltage divider | Built-in voltage divider(1/1000) measures input in the range of ±5V to ±50V (input type "7" only) |
Non-standard input |
Refer to the table of standard range and programmable minimum width of scale Minimum width of scale --- DC voltage: 10mV DC width or mor Thermocouple: K; 250°C width or moreE,J,T; 200°C width or moreR; 800°C width or more Resistance thermometer: 100°C width or more |
Double scale | Measures input with 2 types of scales (each scale is only serial channel) |
Burnout |
Function for detecting disconnection for sensor with thermocouple or resistance thermometer input. Up-scale and down scale burnout on all channels can be programmed (Input type "7" only), parallel operation is not possible |
Chart speed Hour/minute change | 5-speed change, 5,10,20,40,80mm/minute, hour change |
16m chart paper | Maximum length 15.6m |
EL3000 Series Electronic Recorders
Items | Recorder Type | Part Number | Qty | Price |
EL3000 Ink Pad Case |
EL3000/ES600 1 Point Dotting |
22023-039001 81-0056 |
1 Box (5 ea.) | |
EL3000 Ink Pad Case |
EL3000/ES600 2 Points Dotting |
22023-039002 81-0067 |
1 Box (5 ea.) | |
EL3000 Ink Pad Case |
EL3000/ES600 3 Points Dotting |
22023-039003 81-0078 |
1 Box (5 ea.) | |
EL3000 Ink Pad Case |
EL3000/ES600 6 Points Dotting |
22023-039006 81-0089 |
1 Box (5 ea.) | |
Chart Paper - 100mm (Standard) | All Type | EM001 | 1 Box (15 sheets) |
Performance Specification Sheet | Analog Recorder | EL3000 Series | Download |